R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.12
E-3
[SOURce:]TSIGnal:RECT:CREate:NAMed
<waveform filename>
The command generates a rectangular test signal and stores the file on the local hard disk (HD).
The
Create Test Signal
window opens automatically and the signal can be stored as a waveform
file.
This command triggers an event and therefore has no *RST value and no query form.
Example:
TSIG:RECT:CRE:NAM "D:\temp\test.wv"
'writes the generated waveform file to 'test.wv' on the HD.
*RST value
Resolution
Options
SCPI
-
Device-specific
[SOURce:]TSIGnal:RECT:OFFSet
-1 ... +1 FS
The command sets a DC component.to the rectangular test signal.
Example:
TSIG:RECT:OFFS 0.1
'sets an offset of 0.1 FS.
*RST value
Resolution
Options
SCPI
0
Device-specific
[SOURce:]TSIGnal:SINE:SAMPles
4 ... 1000 samples
The command sets the number of sample values required from the rectangular wave per period.
The resulting clock rate must not exceed the maximum ARB clock rate of 300 MHz.
Example:
TSIG:RECT:SAMP 500
'sets the number of samples to 500 for a period.
*RST value
Resolution
Options
SCPI
100
Device-specific
[SOURce:]TSIGnal:SINE:CREate
The command generates a waveform which is output straight away.
This command triggers an event and therefore has no *RST value and no query form.
Example:
TSIG:SINE:CRE
'a signal is generated and output directly.
*RST value
Resolution
Options
SCPI
-
Device-specific