R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.53
E-3
[SOURce:]EQUalizer:MODulator:CATalog?
This command lists all files containing equalizer data tables for the modulator saved on local hard
disk. The default directory is set using command
MMEM:CDIRectory
.
The command is a query command and therefore has no *RST value.
Example:
MMEM:CDIR 'd:/user/Equalizer'
'sets the default directory
EQU:MOD:CAT?
'lists all files containing equalizer data tables for the modulator.
*RST value
Resolution
Options
SCPI
-
-
Device-specific
[SOURce:]EQUalizer:MODulator:DELete
<FILE NAME>
This command deletes an equalizer data table for the modulator from local hard disk.
The command triggers an event and therefore has no query form and no *RST value.
Example:
EQU:MOD:SEL "test_mod"
'selects the file 'test_mod.afq_mfg' to be deleted.
EQU:MOD:DEL "test_mod"
'deletes the file 'test_mod.afq_mfg' from HD.
*RST value
Resolution
Options
SCPI
-
-
Device-specific
[SOURce:]EQUalizer:MODulator:FEMode
ABSolute | RELative
This command determines whether frequency values are absolute or relative with respect to the RF
carrier frequency.
Example:
EQU:MOD:FEM REL
'frequency is given relative to the RF carrier frequency.
*RST value
Resolution
Options
SCPI
ABSolute
Device-specific
[SOURce:]EQUalizer:MODulator:FREQuency
0 ... 1 GHz
This command enters frequency values into the current equalizer data table of the modulator.
Example:
EQU:MOD:FREQ 10000,20000,30000
'enters three requency values into the current data table.
*RST value
Resolution
Options
SCPI
-
1 Hz
Device-specific