R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.36
E-3
Baseband Block - Marker Settings
Command Parameter
Default
unit
Note
[SOURce:]TRIGger:OUTPut:DELay:FIXed ON
|
OFF
[SOURce:]TRIGger:OUTPut<[1]...4>:DELay
0 ... 2^20-1 Samples
S
[SOURce:]TRIGger:OUTPut<[1]...4>:DELay:MAXimum? -
Query
only
[SOURce:]TRIGger:OUTPut<[1]...4>:DELay:MINimum? -
Query
only
[:SOURce]:TRIGger:OUTPut<[1]...4>:LIST -
[SOURce:]TRIGger:OUTPut<[1]...4>:MODE
UNCHanged | RESTart | PULSe |
PATTern | RATio | LIST
[SOURce:]TRIGger:OUTPut<[1]...4>:OFFTime
1 ... max_wavelength -1 samples
S
[SOURce:]TRIGger:OUTPut<[1]...4>:ONTime
1 ... max_wavelength -1 samples
S
[SOURce:]TRIGger:OUTPut<[1]...4>:PATTern
[SOURce:]TRIGger:OUTPut<[1]...4>:PULSe:DIVider 2...2^10
[SOURce:]TRIGger:OUTPut<[1]...4>:PULSe:FREQ?
Query
only
[SOURce:]TRIGger:OUTPut:DELay:FIXed
ON | OFF
The command restricts the marker delay setting range to the dynamic range. In this range the delay
can be set without restarting the marker and signal. If a delay is entered in setting ON but is outside
this range, the maximum possible delay is set and an error message is output.
The numeric suffix in
OUTPut
has no significance for this command, since the setting always affects
every marker.
Example:
TRIG:OUTP:DEL:FIX ON
'restricts the marker signal delay setting range to the dynamic range.
*RST value
Resolution
Options
SCPI
OFF
-
Device-specific
[SOURce:]TRIGger:OUTPut<[1]...4>:DELay
0 ... 2^20-1 Samples
The command defines the delay between the signal on the marker outputs and the start of the
signals, expressed in terms of samples. Command
:TRIGger:OUTPut:DELay:FIXed ON
can be
used to restrict the range of values to the dynamic range, i.e. the range within which a delay of the
marker signals can be set without restarting the marker and signal.
Example:
TRIG:OUTP2:DEL 16
'sets a delay of 16 samples for the signal on connector MARKER 2.
*RST value
Resolution
Options
SCPI
0
1 S
Device-specific