AFQ Commands - Setup Section
1401.3084.32 6.94
E-3
SYSTem:PRESet
The command triggers an instrument reset. It has the same effect as the the *RST command.
The command triggers an event and therefore has no *RST value and no query form.
Example:
SYST:PRES
'all instrument settings (also those that are not currently active) are reset to
their default values.
*RST value
Resolution
Options
SCPI
-
-
Compliant
SYSTem:PROTect<n>[:STATe]
ON | OFF, password
The command activates and deactivates the specified protection level. There are several protection
levels which disable specific service functions (authorized personnel of R&S Service Departments
only). These levels are identified by the suffix under PROTect.
The respective functions are disabled when the protection level is activated. No password is
required for activation. A password must be entered to deactivate the protection level. The password
for the first level is 123456. This protection level can be used to lock-out internal adjustments.
Example:
SYST:PROT1 ON
'activates protection level 1. Internal adjustments are only possible after
deactivating the lock-out.
SYST:PROT1 OFF, 123456
'deactivates protection level 1. Internal adjustments are enabled again.
*RST value
Resolution
Options
SCPI
-
-
Compliant
SYSTem:SERRor?
This command returns a list of all errors existing at the time when the query is started . This list
corresponds to the display on the info page under manual control.
The command is a query command and therefore has no *RST value.
Example:
SYST:SERR?
'queries all errors existing in the error queue.
Response:
"-221, 'Settings conflict', 153, 'Input voltage out of
range'"
'the two returned errors have occurred since the error queue was last queried.
*RST value
Resolution
Options
SCPI
-
-
Device-specific