R&S AFQ100A
Baseband Block
1401.3084.32 4.3
E-3
Baseband Block
The settings for activation and configuration of the external or internal waveform signal can be
accessed in the block diagram via the
Baseband
function block or by selecting
Baseband
in the
Edit
menu.
ARB Dialog
The dialog for setting the Arbitrary Waveform Generator (
ARB
) can be opened in the
Baseband
block
of the main window, clicking the
config...
button and selecting
ARB
.
The
ARB
dialog is used for powering up, selecting the ARB waveform and for configuring a sinusoidal,
rectangular or a continuous test signal.
State - ARB
Switches on and off the ARB modulation.
The output is based on the waveform file that is loaded. The name of
the waveform file is indicated in the header section of the display.
If no waveform file is loaded, the ARB modulation cannot be powered
up.
None
will be displayed next to
Load Waveform
. Additionally a
message is displayed in the info line of the status bar, asking the user
to load a waveform file.
Error message, displayed in the info line: "Settings conflict - no or
empty waveform selected."
Remote-control command:
SOUR:STAT ON