R&S AFQ100A
Baseband Block
1401.3084.32 4.10
E-3
The two sine waves can be configured in the sections Sine 1 and Sine 2.
Periods - Table Sine
Test Signal
Sets the number of periods for the sine waves. The values are set
separately for each component of the test signal, as provided in
section Sine 1 and section Sine 2 of the dialog.
Remote-control command:
SOUR:TSIG:TSIN1:PER 5
SOUR:TSIG:TSIN2:PER 7
Amplitude - Table Sine
Test Signal
Sets the amplitude values of the sine waves. The values are set
separately for each sine wave of the test signal, as provided in section
Sine 1 and section Sine 2 of the dialog.
Remote-control command:
SOUR:TSIG:TSIN1:AMPL 0.3
SOUR:TSIG:TSIN2:AMPL 0.4
Phase - Table Sine
Test Signal
Sets the phase offset values of the sine waves. The values are set
separately for each sine wave of the test signal, as provided in section
Sine 1 and section Sine 2 of the dialog.
Enters the phase shifts of the two sine waves.
Remote-control command:
SOUR:TSIG1:TSIN1:PHAS 360
SOUR:TSIG2:TSIN2:PHAS -360
Res. Frequency -
Table Sine Test Signal
Displays the resulting frequencies of the sine waves. The values of
both components of the test signal are indicated separately in section
Sine 1 and section Sine 2, respectively.
Remote-control command:
SOUR:TSIG1:TSIN1:FREQ?
SOUR:TSIG2:TSIN2:FREQ?
Generate Signal HD -
Table Sine Test Signal
Generates a signal and saves it to a file on HD.
The
Create Test Signal
window opens automatically and the signal
can be stored as a waveform file.
Remote-control command:
SOUR:TSIGl:TSIN:CRE:NAM
Generate Signal RAM -
Table Sine Test Signal
Generates a signal and uses it as output straight away.
Remote-control command:
SOUR:TSIG:TSIN:CRE