R&S AFQ100A
Baseband Block
1401.3084.32 4.28
E-3
Pulse
A regular marker signal (strobe) is generated. The
pulse frequency is defined by entering a divider. The
frequency is derived by dividing the sample rate by
the divider. When
Pulse
is selected an input box for
the divider and the resulting pulse frequency are
displayed (see also
Remote-control command:
SOUR:TRIG:OUTP1:MODE PULS
SOUR:TRIG:OUTP1:PULS:DIV 4
SOUR:TRIG:OUTP1:PULS:FREQ?
Pattern
A marker signal that is defined by a bit pattern is
generated. The pattern has a maximum length of 32
bits and is defined in an input field. When
Pattern
is
selected, an additional button indicating a binary
number appears on the right.
Clicking on this button, the input field for entering the
the pattern opens (see also
Remote-control command:
SOUR:TRIG:OUTP1:MODE PATT
SOUR:TRIG:OUTP1:PATT #11F
Ratio
A regular marker signal that is defined by an on/off
ratio is generated. A period lasts one on and off
cycle.
ON time
ON time
OFF time
OFF time
Start of signal
When
Ratio
is selected the input boxes for entering
the ON/OFF time are displayed (see also
, on page 4.30). The ON time and OFF
time are each expressed as a number of samples.
Remote-control command:
SOUR:TRIG:OUTP1:MODE RAT
SOUR:TRIG:OUTP1:OFFT 30