R&S AFQ100A
AFQ Commands - Signal Generation
1401.3084.32 6.71
E-3
[SOURce:]INPut:BERT:THReshold
0 ... 2.0 V
The command sets the high/low threshold of the BERT inputs
Clock
,
Data
,
Data Enable
and
Restart
. In the case of positive polarity, this threshold determines the point as of which a signal is
high (active) or low (inactive).
Example:
INP:BERT:THR 1 V
'a high/low threshold of 1 volt is set at the BERT inputs. The signal is high
(active) for a signal voltage of 1 volt and higher.
*RST value
Resolution
Options
SCPI
1.00 V
-
Device-specific
TRIGger:BERT:IMMediate
The command triggers a BERT measurement.
This command triggers an event and therefore has no *RST value and no query form.
Example:
TRIG:BERT:IMM
'triggers a BERT measurement.
*RST value
Resolution
Options
SCPI
-
-
Device-specific