Example 2 BS3680 U-Throated Flume
In this example it is required to calculate to BS3680 the flow through a U-
Throated Flume without any hump. Absolute calculation will be used, and
there is a requirement for an alarm to indicate a low flow condition which
will be set to relay 1. The flow rate is to be displayed in cubic meters/hour
and the totaliser is also to record the flow in cubic metres, the resettable
totaliser is to be displayed during RUN.
The distance from the end of the transducer horn (dB Mach 3) to
zero
flow
(
P105
) is 1 metre and
max head
(
P704
) is 0.4 metres,
maximum
flow
(
P705
) which will be calculated by the Ultra unit as 725.171 cubic
metres/hour.
The dimensions of the flume are as follows:
Approach
Channel
diameter
(
Dim “A”
)
P710
= 0.7 m
Throat diameter
(
Dim “B”
)
P711
= 0.5 m
Throat length
(
Dim “C”
)
P712 =
1.0 m
Содержание ULTRA 5
Страница 1: ...ULTRA 5 INSTRUCTION MANUAL ...
Страница 2: ......
Страница 4: ......
Страница 15: ...Page 7 EU Declaration of Conformity Wall Mount ...
Страница 16: ...Page 8 Fascia Mount ...
Страница 17: ...Page 9 Rack and Panel ...
Страница 18: ...Page 10 This page left blank intentionally ...
Страница 21: ...Page 13 The full dimensions of the enclosure are as shown below ...
Страница 23: ...Page 15 The full dimensions of the Fascia mount enclosure are as shown below ...
Страница 25: ...Page 17 Dimensions of the panel enclosure are as shown below VOLTAGE SELECTOR AC POWER FUSE 230 115 ...
Страница 35: ...Page 27 This page left blank intentionally ...
Страница 151: ...Page 143 This page left blank intentionally ...
Страница 240: ...Page 232 This page left blank intentionally ...