120-0313 Rev 04
4–15
Chapter 4 – Customization
Time
[
06:00:00am
,
hh:mm:ss
am/pm ], where:
hh
Two-digit hour value, with leading zero
mm
Two-digit minute value
ss
Two-digit second value
Defines the time of day to trigger automatic calibration with the internal
weight.
NOTE:
Parata recommends that you set the automatic calibration time to a
time outside of operating hours.
NOTE:
Only available in Automatic calibration mode.
Calibration weight
[
0.000g
,
n.nnn
g ]
Defines the external calibration weight.
NOTE:
Only available in Ext. user-def.weight calibration mode.
Weight ID
(External)
[
Off
, On ]
Enables a prompt during External calibration to enter an alphanumeric
identifier for the external weight (e.g., you have more than one
200-gram calibration weight).
When used, the Weight ID appears in the calibration history
(see “Calibration history” on page 1-10).
NOTE:
Only available as a toggle in External calibration mode.
Weight ID
(Ext. user-define.weight)
[
(none)
,
any text string up to 21 characters
]
Defines the alphanumeric identifier for the external, user-defined
calibration weight.
When defined, the Weight ID appears in the calibration history
(see “Calibration history” on page 1-10).
NOTE:
Only available as a text string in Ext. user-def.weight calibration mode.
Table 4-10:
Calibration settings (cont’d)
Setting
Function
Содержание AccuCount II
Страница 8: ...AccuCount II User Guide viii 120 0313 Rev 04...
Страница 24: ...AccuCount II User Guide 1 16 120 0313 Rev 04 Installation notes...
Страница 54: ...AccuCount II User Guide 3 18 120 0313 Rev 04 Operations notes...
Страница 122: ...AccuCount II User Guide 4 68 120 0313 Rev 04 Customization notes...
Страница 145: ...120 0313 Rev 04 5 23 Chapter 5 Maintenance Maintenance notes...
Страница 146: ...AccuCount II User Guide 5 24 120 0313 Rev 04...
Страница 150: ...AccuCount II User Guide 6 4 120 0313 Rev 04 Troubleshooting notes...