Chapter 6 – Troubleshooting
The display flashes continuously during
calibration.
The draft is too strong at the scale location.
• Place the compounding plate on top of the draft
shield while calibrating, to minimize the effects of
the draft.
The scale location is not stable enough.
• Interrupt the calibration using the
On/Off
button
and then relocate the scale.
Use of an imprecise calibration weight (only applies to
external calibration).
• Use the correct calibration weight.
Configuration menu cannot be changed. The password lock is activated.
Symptom
Possible causes and solutions