AccuCount II User Guide
1–12
120-0313
Rev 04
NOTE
The screw hole is on the bottom-right near the front, when the scale is upright.
The scale switches into
Service mode
, whose display is similar to the Weighing application.
8
Slide the scale back onto the surface and level it (see “Leveling” on page 1-8).
9
Press and hold the
Tare
key until the Calibration menu appears.
10 Use the
and
soft keys to select
Calibration
.
11 Press the
Enter
soft key.
The scale begins a special calibration process that is only available in Service mode, during
which it performs three calibrations:
Internal
External
Internal