1.3 Leveling and calibration
The AccuCount II has a high-precision balance which must be leveled and calibrated
for the scale to function properly. When you first install the scale and any time that the
scale is physically moved, perform the leveling and calibration functions in this
section.
NOTE
It is the owner's responsibility to calibrate the AccuCount II.
1.3.1 Leveling
To function properly, the scale must be precisely horizontal. The scale is fitted with
one spirit level and two adjustable feet for level-control. The feet can be used to
compensate for small height differences and/or unevenness in the surface where the
scale is placed. The two feet must be adjusted as follows:
• both of the feet are in contact with the surface
• the air bubble is precisely in the center of the bubble level, as shown below:
NOTE
In order to get exact measurements, the scale
must
be carefully leveled during initial
installation and after each time that it is moved or relocated.
Do not slide or move the scale when it is not in use, as it must be leveled before being
used again.
1.3.2 Calibration
Calibration of the scale must be carried out on initial installation and after each
subsequent relocation. To ensure exact measurements, calibration of the scale every
30 to 60 days is recommended.
Do not move or slide the scale when it is not in use, as it must be recalibrated before
being used again.
Incorrect
Correct
Содержание AccuCount II
Страница 8: ...AccuCount II User Guide viii 120 0313 Rev 04...
Страница 24: ...AccuCount II User Guide 1 16 120 0313 Rev 04 Installation notes...
Страница 54: ...AccuCount II User Guide 3 18 120 0313 Rev 04 Operations notes...
Страница 122: ...AccuCount II User Guide 4 68 120 0313 Rev 04 Customization notes...
Страница 145: ...120 0313 Rev 04 5 23 Chapter 5 Maintenance Maintenance notes...
Страница 146: ...AccuCount II User Guide 5 24 120 0313 Rev 04...
Страница 150: ...AccuCount II User Guide 6 4 120 0313 Rev 04 Troubleshooting notes...