120-0313 Rev 04
3–15
Chapter 3 – Operations
e.
On the verification screen, press
OK
.
The new drug’s piece weight is stored in the user database.
f.
Proceed to step 5.
If the drug
is
in the library, take the following actions:
a.
Manually count the reference quantity (100 pieces or the entire contents of the stock
bottle, if within scale capacity).
b.
Place the pills in the scoop on the scale.
c.
Ensure that the correct quantity is displayed.
If the
incorrect
quantity is displayed, proceed to step 4.
If the
correct
quantity is displayed, proceed to step 5.
4
Drug is in the library,
but
the quantity is not correct:
NOTE
If you are using the entire contents of a stock bottle, the incorrect quantity might be
caused by an incorrect number of pills contained in the stock bottle.
a.
Press the
Recal
soft key.
b.
Press the soft key that corresponds to the number of pills that you counted;
or press the
Var
soft key to enter a different number.
The scale calculates the programmed piece weight.
c.
On the verification screen, press
OK
.
The new programmed piece weight is now stored in the library.
d.
Proceed to step 5.
5
Drug is in the library, and the quantity is correct:
a.
Place the pills back into the stock bottle.
b.
Count the prescription quantity.
c.
Mark the stock bottle with its remaining pill quantity.
Содержание AccuCount II
Страница 8: ...AccuCount II User Guide viii 120 0313 Rev 04...
Страница 24: ...AccuCount II User Guide 1 16 120 0313 Rev 04 Installation notes...
Страница 54: ...AccuCount II User Guide 3 18 120 0313 Rev 04 Operations notes...
Страница 122: ...AccuCount II User Guide 4 68 120 0313 Rev 04 Customization notes...
Страница 145: ...120 0313 Rev 04 5 23 Chapter 5 Maintenance Maintenance notes...
Страница 146: ...AccuCount II User Guide 5 24 120 0313 Rev 04...
Страница 150: ...AccuCount II User Guide 6 4 120 0313 Rev 04 Troubleshooting notes...