Chapter 4 – Customization
Table 4-22 provides an overview of each Minimal sample weight option with all
possible values (default settings are shown in
bold text
).
4.2.8 Properties
Select the
Properties
configuration item to display all software settings for the scale’s
configuration in an info screen.
See “Info screens” on page 2-9 for more information about using this screen.
Table 4-22:
Minimal sample weight settings
Setting
Function
Tare range 1|2|3 (Max.)
[
0.000 g
,
nnn.nnn
g ], where:
nnn.nnn
Any decimal value between 0.000 and 620.000,
in grams
Sets the tare range 1, 2, or 3.
Tare range is the capacity range within which a scale can still
display a net weight after taring.
Min. sample weight 1|2|3
[
0.000 g
,
nnn.nnn
g ], where:
nnn.nnn
Any decimal value between 0.000 and 620.000,
in grams
Sets the minimal sample weight 1, 2, or 3.
Minimum sample weight is the minimum sample quantity required
by the scale to obtain a reliable result, taking into account factors
such as measurement uncertainty.