AccuCount II User Guide
1–10
120-0313
Rev 04
1.3.2.b
Automatic calibration
You can set the AccuCount II to calibrate automatically at a particular time of day, when
the ambient temperature changes by a set amount, or both.
Automatic calibration performs an internal calibration based on those settings
(see “Calibration” on page 4-13 for more information).
NOTE
Parata recommends that you set the automatic calibration time to a time outside of
operating hours.
1.3.2.c
Calibration history
The AccuCount II maintains a record of every calibration that it conducts, be the
calibration automatic or requested by a user. Each record includes the following
details:
• Calibration type (see “Calibration” on page 4-13)
• Date and time
• Ambient temperature
• User logged into the scale at the time of calibration
• Calibration results (e.g.,
Calibration OK
)
To check the calibration history
1
Select the Weighing application from the Home screen using the soft keys (see “Applications
menu” on page 3-3).
2
Press-and-hold the
Tare
key until the Calibration menu appears.
3
Use the
and
soft keys to select
Calibration history
.
4
Press the
Enter
soft key.
The calibration history info screen appears (scrolled down to show a record, below).
See “Info screens” on page 2-9 for more information about using this screen.
Содержание AccuCount II
Страница 8: ...AccuCount II User Guide viii 120 0313 Rev 04...
Страница 24: ...AccuCount II User Guide 1 16 120 0313 Rev 04 Installation notes...
Страница 54: ...AccuCount II User Guide 3 18 120 0313 Rev 04 Operations notes...
Страница 122: ...AccuCount II User Guide 4 68 120 0313 Rev 04 Customization notes...
Страница 145: ...120 0313 Rev 04 5 23 Chapter 5 Maintenance Maintenance notes...
Страница 146: ...AccuCount II User Guide 5 24 120 0313 Rev 04...
Страница 150: ...AccuCount II User Guide 6 4 120 0313 Rev 04 Troubleshooting notes...