![Oxford Instruments OpAL Скачать руководство пользователя страница 84](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943084.webp)
OpAL
Oxford Instruments Plasma Technology
System Manual
3.7.2
Upper chamber configuration for thermal ALD
The upper chamber configuration for thermal ALD is shown in Fig 3-24 (based on
drawing G-MA71A29920).
This configuration comprises the following major components:
Chamber top cover incorporating a process gas inlet.
Process chamber top – forms the upper half of the process chamber and when lowered
onto the lower chamber provides a vacuum-tight process chamber. Sealing is provided by
integral O rings. Apertures are provided for the following:
Two blanked view ports – each can be used as a view port or to mount an end point
detection device.
Process gas line inlet.
Process gas valve
(if fitted)
Chamber top cover
(thermal)
O ring
O ring
Clamping
components
Clamping
components
Blank view port
components
Gas port assembly
Liner disk
Process
chamber top
Centering ring
Fig 3-24: Upper chamber configuration for thermal ALD