OpAL
Oxford Instruments Plasma Technology
System Manual
3.4.4
Heating
The management of condensation and decomposition in
OpAL
is achieved by dedicated
temperature control of individual components. Up to 16 temperature zones are available
as shown in Fig 3-16 and Table 3-1. Temperature is controlled using PID.
Zone 1
Zone 2
Services panel
Gate
Valve
ICP
Gate Valve
Zone 4
External pumping
4 x 0.5 m lengths
Zone 3
Precursor 1
Reserved for water
Zone 5
Precursor 2
Zones 6, 7, 8
Precursor 3
Precursor 4
Zones 9, 10,11
Zones 12,13,14
Fig 3-16: Temperature zones
Zone
Name
Temp Range (deg C)
Power (Watts)
Snap switch
1
Upper Chamber
20 - 180
2000
200 ± 10
2
Lower Chamber
20 - 180
2000
200 ± 10
3
Internal pumping
20 - 120
75
160 ± 10
4
External pumping
20 -120
4 x75
160 ± 10
5
P1 –Water Z3 (dual valve)
20 - 200
84
220 ± 10
6
P2Z1 (source pot)
20 - 200
87
220 ± 10
7
P2Z2(intermediate)
20 - 200
74
220 ± 10
8
P2Z3(dual valve)
20 - 200
84
220 ± 10
9
P3Z1(source pot)
20 - 200
87
220 ± 10
10
P3Z2(intermediate)
20 - 200
74
220 ± 10
11
P3Z3(dual valve)
20 - 200
84
220 ± 10
12
P4Z1(source pot)
20 - 200
87
220 ± 10
13
P4Z2(intermediate)
20 - 200
74
220 ± 10
14
P4Z3(dual valve)
20 - 200
84
220 ± 10
15
ICP Gate Valve
20 - 120
162
160 ± 10
16
Spare
20 - 200
220 ± 10
Table 3-1: Temperature zones
Description
Issue 1: August 07
Page 3-16 of 46
Printed: 07 August 2009 07:42