![Oxford Instruments OpAL Скачать руководство пользователя страница 309](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943309.webp)
System Manual
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó==
^ää=lfmq=póëíÉãë
Drive C1 negative to the stop position; LED 102 lights; stop point in the range 000 -
050
Drive C2 positive to the stop position; LED 1 lights; stop point in the range 950-999
Drive C2 negative to the stop position; LED 2 lights; stop point in the range 000 - 050
If these aren’t working as stated, refer to section 5 or OIPT Work Instruction No. 39
8)
Set LK2 and LK 102 to
position ‘b’, and then
turn RV1 and RV101 fully
clockwise until they
begin to click (see Fig 6
RV1
LK2
JP3
LK102
RV101
Fig 6: Component locations
9) Fit all covers to the AMU ensuring that they are securely fitted, and then connect the
RF generator to the matching unit. Evacuate the process chamber and turn on a low
power process.
For a Plasmalab system, a suitable process would be:
RF Generator output:
50 W
Pressure:
50 mTorr (RIE), 1 Torr (PECVD), 0.1 Torr (PE)
Gas:
20-100 sccm air, nitrogen or argon.
For an ion beam system, a suitable process would be:
RF generator output: 150 W (3 cm to 5 cm diameter)
300 W (15 cm to 20 cm diameter)
Gas:
10 sccm Argon
It may be necessary to use a gas burst to start the plasma, and it may be
necessary to start the neutralizer, if fitted.
OIPT Automatch Unit
Printed: 16-Nov-07, 9:26
Page 11 of 20
Issue 6: September 05
Содержание OpAL
Страница 14: ......
Страница 229: ...OpAL Open Load Atomic Layer Thermal Deposition System Installation Data Typical OpAL thermal system Issue 3 October ...
Страница 299: ...bèìáéãÉåí j åì ä ää lfmq póëíÉãë RF Automatic Matching Unit Issue 6 September 2005 Page 1 of 20 ...
Страница 319: ...ALD Ozone Delivery System Ozone Delivery System for FlexAL and OpAL Systems Equipment Manual Issue 1 July 2010 ...