System Manual
=
lñÑçêÇ=fåëíêìãÉåíë=mä~ëã~=qÉÅÜåçäçÖó=
lé^i
RF
AUTOMATCH
panel
CAPACITOR 1
field
Enter the required position for variable AMU Capacitor 1. The position can
be set between 0%, minimum capacitance, and 100%, maximum
capacitance. The capacitor position read back is displayed.
CAPACITOR 2
field
Enter the required position for variable AMU Capacitor 2. The position can
be set between 0%, minimum capacitance, and 100%, maximum
capacitance. The capacitor position read back is displayed.
AUTO
button
Select to enable the AMU to tune automatically when the RF generator is
switched on. When the RF generator is switched off, the capacitors return
to the park position.
MANUAL
button
Select to enable the AMU to move the capacitors to the values defined in
the CAPACITOR 1 and CAPACITOR 2 fields; the capacitors will remain in
these positions.
HOLD
button
Select to enable the AMU to tune automatically when the RF generator is
switched on. When the RF generator is switched off the capacitors remain
at the last position.
ICP Gate
Valve
panel
Enter a ‘1’ to open the valve or ‘0’ to close it. An indicator displays current
status.
The ICP gate valve should be open when running process gases and when
running a plasma step, it should be closed during a precursor process step
to stop coating of the ICP tube with precursor.
Operating Instructions
Printed: 06 November 2007 09:42
Page 5-49 of 60
Generic (OpAL Plasma) Issue 1: Oct. 07