píÉé=Q=Emçí=éêÉé~ê~íáçå=Ñçê=ÄìÄÄäáåÖF=
1) Purge bubbler lines.; see Fig 2-5.
2) Turn on argon purge and argon bubbler MFCs and pump out any residuals in the lines..
3)
Do not turn on the manual valves yet!
To chamber
Heated TEMAH pot pre-
fitted with manual valves.
Single
ALD valve
jacket heated
to 200 °C
Drip tray
Extraction port
1 m
3
/h
Precursor cabinet
N
2
purge
Purge
Dose
Bypass
Ar purge
500 sccm
Manual valve
Air driven nupro valve –manual control
Fast high temperature ALD valve
Ar bubbler
500 sccm
Switching
E
Fig 2-5: Step 4 (Pot preparation for bubbling)
OKNKNKR
píÉé=R=Emçí=éêÉé~ê~íáçå=Ñçê=ÄìÄÄäáåÖF=
1) Repeat Steps 3 and 4.
2) Carry out a leak check.
3)
Do not turn on the manual valves yet!
Precursor Source Pot
Printed: 7-Oct-07, 12:13
Page 7 of 20
Issue 1: September 07