![Oxford Instruments OpAL Скачать руководство пользователя страница 283](http://html.mh-extra.com/html/oxford-instruments/opal/opal_system-manual_1664943283.webp)
System Manual
Oxford Instruments Plasma Technology
OIPT Systems
DEDICATED
HEAT EXCHANGER
OR
CHILLER
PLASMALAB
OR
IONFAB
SYSTEM
ISOLATION
VALVES
10-MICRON METAL MESH FILTER
MAXIMUM PRESSURE DROP AT
RATED FLOW RATE 0.15 BAR (2.2 PSI)
TOTAL WATER FLOW
METER OF APPROPRIATE
RANGE
(FITTED WITH AN
EXTERNAL BYPASS
HAVING A CAPACITY OF
100% FLOW)
Fig 2.1: Recommended basic recirculation installation
2.1.1
Mandatory Specifications for recirculated water systems
CAUTION
If clear (i.e. transparent) tubing is exposed to sunlight, algal growth can develop,
which can restrict coolant flow.
It is MANDATORY that clear tubing is not used in any part of the cooling system. OIPT
recommends the use of either black or dark green tubing.
CAUTION
It is the customer's responsibility not to exceed a pressure of 4.2 bar, or other limit
that has been set for the system. Exceeding this safe pressure may cause irreparable
damage to system components.
The water must be kept warm enough to prevent condensation on chamber surfaces and
outside system components. This applies to those parts of the system inside the clean
room and those parts in a service area. Condensation can damage components such as
RF power supplies, ferrofluidic seals and automatch units. Any damage so caused cannot
be covered by the system warranty.
Pressure:
Adjustable 0.7 to 4.2 bar (10 to 60 psi). Chiller / heat exchanger to
be fitted with a bypass having a capacity of 100% of rated flow.
Temperature range:
See system installation data sheets.
Minimum flows:
See system installation data sheets.
Cooling capacity:
See system installation data sheets.
Coolant:
Use either Thermal G (94-G-WTR-SUN-909) or Thermal H5 (84-G-
WTR-SUN-903) depending on the application. See sub-section 2.1
for the warranty impact of not using these products.
Filtration:
10 micron metal mesh water filter. Maximum pressure drop 0.15
bar (2.2 psi) at rated flow. For example, filter element Balston SMC-
100-12-10.
Services Specifications
Printed: 6-Jan-10, 8:53
Page 7 of 22