lé^i
a)
Switch the RF power generator(s) OFF.
b)
Switch the pumps OFF.
c)
Close the APC and normally-closed vacuum valves.
d)
Switch the process gases OFF.
After clicking on
Yes
, the system should be restarted using the procedure given in sub-section
Clicking on the
No
button in the Automatic Process Shutdown dialogue will limit the abort to
the actions already done, i.e. stop all automatic recipes. Further processing operations
(automatic or manual) can then be carried out.
RKQKR
póëíÉã=ëÜìíJÇçïå=éêçÅÉÇìêÉ=
This procedure assumes that the process run is finished and that the system is to be shut down
for maintenance or repair.
1)
Ensure that the system has been vented, and all pumps are switched off.
2)
Exit from the PC 2000 software by clicking on the
System
button, then on the
Exit
option.
3)
Exit from Windows.
4)
Turn the PC controller OFF.
5)
At the Main Console, press the
OFF
button (coloured red).
6)
Switch off and lock off the wall-mounted safety isolation box.
7)
Turn off the compressed air supply.
8)
Ensure that all heated components have cooled to ambient temperature, then turn
off the cooling water.
Operating Instructions
Printed: 06 November 2007 09:42
Page 5-7 of 60
Generic (OpAL Plasma) Issue 1: Oct. 07