3.4 Using the Continuous Function
KP3000S
51
3.4.9 Using the Measurement Function
The measured values are displayed in the measured value area (see
5.1
). The main measurement
functions of this product are shown in Table 3-11. For details of the display items of the measured
value, see
5.1.2
.
Table 3-11 Main Measurement Functions
Item Description
Unit
V
Output voltage effective
value
The effective (RMS) value of the output
voltage on the output terminal.
Vrms
Vavg
Output voltage average
value
The average DC value of the output
voltage on the output terminal.
V
Vmax Output
voltage
maximum
peak value
The maximum peak value of the output
voltage on the output terminal.
Vpk
Vmin Output
voltage
minimum
peak value
The minimum peak value of the output
voltage on the output terminal.
Vpk
I
Output RMS current
The effective (RMS) value of the output
current. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Arms
Iavg
Output current average
value
The average DC value of the output
current.
A
Imax Output
current
maximum
peak value
The maximum peak value of the output
current. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Apk
Imin
The minimum peak value
of the output current
The minimum peak value of the output
current. When the line voltage display is
L1-L2/L2-L3/L3-L1, the phase current of
the L1/L2/L3 phase is displayed
respectively.
Apk
Ipk-Hold Output current peak-hold
value
The hold value of Imax or Imin, either of
which the value is bigger. When the line
voltage display is L1-L2/L2-L3/L3-L1,
the phase current of the L1/L2/L3 phase
is displayed respectively.
Apk
P
Active power
The active power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
W
S
Apparent power
The apparent power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
VA
Q
Reactive power
The reactive power provided to the load.
The power of the phase is displayed when
in the phase voltage display. The total
power of all phases is displayed when in
the line voltage display.
var
Содержание KP3000S
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Страница 4: ......
Страница 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Страница 22: ......
Страница 42: ......
Страница 78: ......
Страница 210: ......
Страница 211: ...KP3000S 193 5 Description of Screen and Menu 5 1 Screen Configuration 194 5 2 Menu Composition 198...
Страница 220: ......
Страница 221: ...KP3000S 203 6 Remote Control 6 1 Communication Interface 204 6 2 Remote Local State Switching 210...
Страница 233: ...KP3000S 215 8 Troubleshooting 8 1 Error Messages and Error Handling 216 8 2 When a Failure is Suspected 223...
Страница 252: ......
Страница 286: ......
Страница 288: ......
Страница 289: ......
Страница 290: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...