10.26 External Control I/O
KP3000S
257
10.26 External Control I/O
When you enabled the external control from the menu, you can control this product by using the
external signal (or non voltage contact). The state output is always on. The detection and state
switching is done at 10 ms-cycle. If the Remote state is achieved by the external interface, a control
input is ignored.
Item Description
Control
input
External control input
Enabled or Disabled
Input level
TTL
Nondestructive maximum input
+10 V /
5 V
Input impedance
Pull-up to +5 V at 47 k
Control Output Off
Falling Off
Output On
Falling On
Sequence start/resume
Falling Start
Stop of sequence
Falling Stop
Sequence is in the Hold status
Falling Hold
Sequence branch 1
Falling Branch start
Sequence branch 2
Falling Branch start
Memory recall (+ compile)
Falling Recall
Memory specification 1
Specify 0 to 3
(Equivalent to memory 1 to 4,
respectively)
Memory specification 2
Clear the current peak-hold value
Falling Clear
State
output
Output level
TTL
Output impedance
220
Polarity
Polarity reverse (collective) is available
only for the state with *
Status
Power On/Off
Low: Off, High: On
Output On/Off*
Low: On, High: Off
Factory
default
(negative)
Protection operation*
Low: Active, High: None
Limiter operation*
Low: Active, High: None
AGC/Autocal setting state*
Low: On, High: Off
Software busy*
Low: Busy, High: Ready
Output range
Low: 200 V, High: 100 V
Sequence operation
Step sync 1
High level or Low level
Sequence operation
Step sync 2
Trigger positive
or negative
Terminal D-sub
25-pin
multi-connector
(rear panel, female, M2.6 screw)
Note: Input TTL level (Low: 0.8 V or lower, High: 2.6 V or higher)
Output TTL level (Low: 0.4 V or lower, High: 2.7 V or higher)
All means the chassis potential.
Note: The memory recall input of the control input recalls the setting memory for the Normal
(Continuous), the Sequence Memory for the Sequence, and the Simulation Memory for the
Simulation. For the Sequence and Simulation, the compile data is also included.
Note: Sequence start and stop of the control input are effective for the Simulation as well. Also,
step sync output 1 and 2 of the control output are effective for the Simulation as well.
Содержание KP3000S
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Страница 4: ......
Страница 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Страница 22: ......
Страница 42: ......
Страница 78: ......
Страница 210: ......
Страница 211: ...KP3000S 193 5 Description of Screen and Menu 5 1 Screen Configuration 194 5 2 Menu Composition 198...
Страница 220: ......
Страница 221: ...KP3000S 203 6 Remote Control 6 1 Communication Interface 204 6 2 Remote Local State Switching 210...
Страница 233: ...KP3000S 215 8 Troubleshooting 8 1 Error Messages and Error Handling 216 8 2 When a Failure is Suspected 223...
Страница 252: ......
Страница 286: ......
Страница 288: ......
Страница 289: ......
Страница 290: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...