10.
Specifications
PROGRAMMABLE AC/DC POWER SOURCE
248
10.15 Remote Sensing
This switches the voltage used for measurement. When the remote sensing is on, the sensing input
terminal voltage is used. When it is off, the output terminal voltage is used.
By combining with AGC or Autocal, a voltage drop due to wiring to the load can be compensated.
When the remote sensing is on, the output voltage detection point corrected by the AGC or Autocal
function is switched to the sensing input terminal. When AGC or Autocal is off, only the detection
voltage used for measurement display is switched.
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected.
Measurement
voltage, power, power factor
AGC/Autocal
Off On
On
Use the sensing input terminal voltage
Not active
Active
Off
Use the output terminal voltage
Not active (factory
default)
Active
10.16 AGC
When the AGC (Automatic Gain Control) is on, the detection point voltage is always measured, and
the output voltage is continuously corrected so that its effective value is equal to the output voltage
setting value. The fluctuation of the detection point voltage can be suppressed even when the load is
fluctuated. The detection point can be switched between the sensing input terminal (remote sensing
on) and the output terminal (remote sensing off).
Effective only for AC-INT, AC-VCA, AC-SYNC, DC-INT, and DC-VCA and when the waveform
is sine wave or DC. It is turned off when the Sequence or Simulation is selected. It cannot be
selected when the Autocal is set to on.
KP3000S
Response
time
Within 100 ms (typ.)
(DC/50 Hz/60 Hz, at the rated output voltage)
Operation
range
The output voltage setting is 8 V or higher
Calibration
range
Within
10% (difference between the output voltage and measured value)
The output voltage should be within the allowed voltage setting range of the product.
Accuracy
Within
0.5 V /
1.0 V
(in the case of DC, 40 Hz to 550 Hz, 50 V or higher output voltage,
resistance load, the output current is the maximum current or less)
Содержание KP3000S
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Страница 4: ......
Страница 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Страница 22: ......
Страница 42: ......
Страница 78: ......
Страница 210: ......
Страница 211: ...KP3000S 193 5 Description of Screen and Menu 5 1 Screen Configuration 194 5 2 Menu Composition 198...
Страница 220: ......
Страница 221: ...KP3000S 203 6 Remote Control 6 1 Communication Interface 204 6 2 Remote Local State Switching 210...
Страница 233: ...KP3000S 215 8 Troubleshooting 8 1 Error Messages and Error Handling 216 8 2 When a Failure is Suspected 223...
Страница 252: ......
Страница 286: ......
Страница 288: ......
Страница 289: ......
Страница 290: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...