8. Troubleshooting
PROGRAMMABLE AC/DC POWER SOURCE
226
Table 8-3 Actions When a Failure is Suspected (Continued)
Category Symptom
Possible
Cause Necessary
Actions
Limiter/setti
ng range
limitation-
related
problem
The output
current
peak-hold value
(measured) is
larger than the
peak current
limiter setting
value.
The output current
overshoot occurred for
the limit value.
While the peak current limiter is
activated, the overshoot may
occur depending on the load.
4.1.1
Remote
sensing
problem
The Remote
sensing function
cannot be turned
on.
The output mode is
incorrect.
The Remote sensing function can
be set to on in the AC-INT,
AC-VCA, AC-SYNC, DC-INT,
and DC-VCA modes. However,
for the AC mode, it can be set to
on only when the waveform is a
sine wave. It is turned off when
the Sequence or Simulation
function is selected.
4.11
The waveform setting is
incorrect.
The mode is not the
continuous output mode.
Problem of
AGC
function
The AGC
function cannot
be turned on.
The output mode is
incorrect.
The AGC function can be set to
on in the AC-INT, AC-VCA,
AC-SYNC, DC-INT, and
DC-VCA modes. However, for
the AC mode, it can be set to on
only when the waveform is a sine
wave. It is turned off when the
Sequence or Simulation function
is selected.
4.12
The waveform setting is
incorrect.
The mode is not the
continuous output mode.
The Autocal function is
on.
When the Autocal function is on,
the AGC function cannot be used.
Turn off the Autocal function.
4.13
Problem of
Autocal
function
The Autocal
function cannot
be turned on.
The output mode is
incorrect.
The Autocal function can be set to
on in the AC-INT, AC-VCA,
AC-SYNC, DC-INT, and
DC-VCA modes. However, for
the AC mode, it can be set to on
only when the waveform is a sine
wave. It is turned off when the
Sequence or Simulation function
is selected.
4.13
The waveform setting is
incorrect.
The mode is not the
continuous output mode.
The AGC function is on. When the AGC function is on, the
Autocal function cannot be used.
Turn the AGC function off.
4.12
Problem of
remote
control
Refer to the Instruction Manual (Remote Control) available from our website or
an optional Control Software CD-ROM.
Problem of
external
control
function
The product
cannot be
controlled by the
external control
function.
The external control
function is disabled.
Enable the external control
function.
4.18
The Remote state is set
by the communication
interface.
The external control function is
disabled in the Remote state.
Содержание KP3000S
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Страница 4: ......
Страница 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Страница 22: ......
Страница 42: ......
Страница 78: ......
Страница 210: ......
Страница 211: ...KP3000S 193 5 Description of Screen and Menu 5 1 Screen Configuration 194 5 2 Menu Composition 198...
Страница 220: ......
Страница 221: ...KP3000S 203 6 Remote Control 6 1 Communication Interface 204 6 2 Remote Local State Switching 210...
Страница 233: ...KP3000S 215 8 Troubleshooting 8 1 Error Messages and Error Handling 216 8 2 When a Failure is Suspected 223...
Страница 252: ......
Страница 286: ......
Страница 288: ......
Страница 289: ......
Страница 290: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...