4.17 Setting Voltage using External DC Input Signal
KP3000S
167
2.
Set the output DC voltage in the [DCV] item.
4.17 Setting Voltage using External DC Input Signal
In the AC-VCA mode, you can use the external DC input signal to set the AC output voltage of the
internal signal source. In the DC-VCA mode, you can use the external DC input signal to set the DC
output voltage of the internal signal source. The external DC signal is input to this product through
the external signal input connector. When the signal source is VCA, the specifications except the
output voltage setting are same as when the signal source is INT.
--------
Notes
------------------------------------------------------------------------------------------------------------
For VCA, the external signal input is detected with the interval of about 100 ms.
-----------------------------------------------------------------------------------------------------------------------------
4.17.1 Using External Signal in AC-VCA Mode
In the AC-VCA mode, the peak value of the output AC voltage (ACV) is set according to the
following figure, by using the gain setting value and the voltage of the external DC signal.
ACV (Vop) = Gain x External DC signal voltage (V)
For the specifications of the gain setting range and the external DC signal input voltage range, see
10.22.2
.
--------
Notes
------------------------------------------------------------------------------------------------------------
The gain setting value is retained for each of 100 V/200 V ranges and for each of AC/DC
modes.
Within the same range and the same mode, the same value is retained for the gain setting value,
regardless of the combination of the signal sources (VCA, EXT, ADD).
For the polyphase system, the ACV setting using VCA is common to all phases.
When the switching from AC-VCA mode to ACDC mode, the signal source is forcibly set to
INT.
-----------------------------------------------------------------------------------------------------------------------------
Содержание KP3000S
Страница 1: ...PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL NF Corporation...
Страница 2: ......
Страница 3: ...DA00030908 004 PROGRAMMABLE AC DC POWER SOURCE KP3000S INSTRUCTION MANUAL...
Страница 4: ......
Страница 19: ...KP3000S 1 1 Outline 1 1 Overview 2 1 2 Polyphase System Configuration 2 1 3 Features 2...
Страница 22: ......
Страница 42: ......
Страница 78: ......
Страница 210: ......
Страница 211: ...KP3000S 193 5 Description of Screen and Menu 5 1 Screen Configuration 194 5 2 Menu Composition 198...
Страница 220: ......
Страница 221: ...KP3000S 203 6 Remote Control 6 1 Communication Interface 204 6 2 Remote Local State Switching 210...
Страница 233: ...KP3000S 215 8 Troubleshooting 8 1 Error Messages and Error Handling 216 8 2 When a Failure is Suspected 223...
Страница 252: ......
Страница 286: ......
Страница 288: ......
Страница 289: ......
Страница 290: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima Higashi Kohoku ku Yokohama 223 8508 JAPAN...