![Mobius Bionics LUKE Arm Скачать руководство пользователя страница 96](http://html1.mh-extra.com/html/mobius-bionics/luke-arm/luke-arm_user-manual_1815915096.webp)
User Inputs
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
49
Places Used
Inertial Measurement Units (IMUs) are installed on top of client’s shoes using a shoe
clip or custom fabricated strap or pocket. For detailed information on installing and
setting up the IMUs see
.
Arm Controls
IMUs are used to control arm motion and certain switching functions other than
mode select. See
How They Are Used
If you intend to use IMUs, the first task you need to perform is to examine the
client’s feet and decide if they can be used for viable control inputs. The client has to
have the physical mobility to move their feet with enough dexterity to operate the
IMU.
The IMU designated for motion controls uses all four directions: anterior (toe),
posterior (heel), medial (inside), and lateral (outside). You and the client determine
which of these four motions are used to perform a specific arm motion. You
configure this using the Prosthetist Interface (PI). When configuring the IMUs using
the Prosthetist Interface you can adjust the output gains to vary the speed of the
arm’s motion.
The Prosthetist Interface allows you to configure up to two (2) sets of IMUs. The
client is able to switch between sets. This is useful when an IMU battery runs low;
the client can switch to the second set of IMUs while charging the first set. For
instructions on switching between IMUs, see
Restrictions
The IMUs CANNOT be used to control Switching (Mode Select) of the
arm
. The system will not allow Mode Select to be assigned to an IMU input. This is
to eliminate the risk that the client can activate or deactivate the arm inadvertently
through the simple act of walking.
The arm supports up to four IMU/ACI modules at a time in multiple
combinations, with a maximum of 2 IMUs. For example, you may set up and
configure the arm to support two IMU modules and two wired ACI modules or
four wired ACI modules and no IMUs.
Up to two pairs of IMUs can be assigned to an arm system, however only one
pair may be used at a time. See
for more information about