![Mobius Bionics LUKE Arm Скачать руководство пользователя страница 157](http://html1.mh-extra.com/html/mobius-bionics/luke-arm/luke-arm_user-manual_1815915157.webp)
242
Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document -- Rev. 1.5
Appendix B: Technical Specifications
:
:
Table 35. Operating Environmental Range
Parameter
Range
Operating Temperature
•
10 °C to 40 °C (50 °F to 104 °F) with no degradation in
performance
•
-10 °C to 50 °C (14 °F to 122 °F) with reduced arm speed
and/or load capacity
Humidity
15% to 93% (non-condensing)
Pressure
700 hPa to 1060 hPa
Arm and body worn Mobius
Bionics supplied accessories IP
rating
IP52
IMU IP Rating
IP57
When using the arm continuously in a hot environment (40 °C, 104 °F) and
while charging the internal battery, portions of the forearm and upper arm (if
applicable) could reach temperatures of 54 °C - 57 °C (130 °F - 135 °F), when
evaluated as directed in IEC 60601-1: 2005-12.
Table 36. Transport and Storage Environmental Range
Parameter
Range
Storage Temperature (excluding
battery)
-25
°C
to 70
°C
(-13
°F
to 158
°F
)
Humidity
15% to 93% (non-condensing)
Pressure
700 hPa to 1060 hPa