Chapter 11: Configuring the Arm Using the Prosthetist Interface
Figure 128. Example Configure Action Screen - Toggle Grip Select
If you remove a grip in Toggle Grip Select while that grip is currently selected,
that grip will remain selected until you toggle out of the grip.
Table 26. Configure Actions Settings Glossary
Setting
Description
Single Site
Configures both positive and negative direction
motion in a joint with a single channel that provides
for both positive and negative signal levels, such as a
linear transducer.
Dual Site
Configures both a positive and negative direction
motion in a joint using two separate channels each
with unique gains and thresholds.
IMU
Configures an existing IMU channel for both the
positive and negative direction motion in a joint.
Input Channel
An input signal source, connected to a specific type of
sensor the client uses to signal control intent.