Chapter 14: Maintaining and Troubleshooting the Arm
Arm Mode LED Blinking
Walk Detect
1.
Stop walking
2.
Put your feet flat on the ground
3.
Check that the Arm Mode LED has
stopped blinking
IMU tilted too far
1.
Put the arm into Standby Mode
2.
Check and secure the IMUs on your
feet
3.
Put your feet flat on the ground
4.
Take the arm out of Standby Mode
Thresholds or gains need
adjustment
1.
Put the arm into Standby Mode
2.
Ask the client to be at a resting position
3.
Use PI to check that all non-IMU inputs
are below the threshold
4.
Adjust the input device thresholds or
gains as necessary
5.
Update the arm
6.
Take the arm out of Standby Mode
Input device not working
1.
Power the arm off
2.
Replace the input device
3.
Power the arm on
4.
Use PI to check the thresholds and
gains for the new input device and
adjust if necessary
Table 30. Troubleshooting - Wrist Display and System Faults
Problem
Cause
Solution