Troubleshooting the Arm
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
231
Table 31. Troubleshooting - Arm Function
Problem
Cause
Solution
Arm moving without command
IMU zero position
changed
1.
Put the arm into Standby Mode
2.
Check and secure the IMUs on your
feet.
3.
Put your feet flat on the ground
4.
Take the arm out of Standby Mode
Sweat near EMGs
1.
Put the arm into Standby Mode
2.
Wipe the sweat from the EMG
electrode and skin with a dry cloth
3.
Take the arm out of Standby Mode
Ham radios or walkie
talkies are affecting the
arm
1.
Put the arm into Standby Mode
2.
Move the arm at least 0.5 m (20
inches) away from any ham radios or
walkie talkies
3.
Take the arm out of Standby Mode
Arm not moving
Arm is in Standby Mode
Put the arm into Hand Mode
Arm is off
1.
Shake the IMUs to wake them
2.
Power the arm on
Arm is faulted
See
Table 30, Troubleshooting - Wrist
Input device not
connected
1.
Power the arm off
2.
Check and tighten all connections at
the ACI
3.
Power the arm on
Cannot change Modes
Input device not
connected
1.
Power the arm off
2.
Check and tighten all connections at
the ACI
3.
Power the arm on
Input device not working
1.
Power the arm off
2.
Replace the input device
3.
Power the arm on
4.
Use PI to check the thresholds and
gains for the new input device and
adjust if necessary