Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document -- Rev. 1.5
Chapter 11: Configuring the Arm Using the Prosthetist Interface
Select the Arm Assembly
From the
Select Arm Assembly
drop down list, select the correct arm
configuration for the client.
The possible values are:
•
Shoulder Configuration
•
Humeral Configuration
•
Radial Configuration
When you select an arm assembly, a picture of the arm’s configuration is displayed
in the arm illustration box on the PI main screen each time the client file is opened.
Zero the Shoulder (if fitting an SC arm)
After initially mounting the SC arm to the socket, see
, you
must configure the shoulder to the neutral position.
To zero the shoulder or configure the arm’s neutral position, perform the following
steps:
1.
Have the client don the Arm System and stand up straight.
2.
Click the
Zero Shoulder
button.
The Configure the Shoulder’s Neutral Position screen opens. See