![Mobius Bionics LUKE Arm Скачать руководство пользователя страница 112](http://html1.mh-extra.com/html/mobius-bionics/luke-arm/luke-arm_user-manual_1815915112.webp)
168
Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document -- Rev. 1.5
Chapter 11: Configuring the Arm Using the Prosthetist Interface
Figure 102. Power Save Mode
2.
Enter the new delay value (in seconds), and click
Update Arm with New
Delay
.
3.
Click
Done
.
Real Time Clock
Selecting the Real Time Clock function sends the computer’s clock settings to the
arm system. This sets the arm's clock. The arm's clock is set at the factory. The
arm’s clock is used to put a time stamp on the event log.
The factory sets the arm's clock to Eastern time, and the clock does not
automatically adjust for daylight savings time.
To get or set the Real Time Clock perform the following steps:
1.
From the
Settings
menu select
Real Time Clock
.
The Real Time Clock screen opens. See
Figure 103. Real Time Clock
2.
To get the arm’s current clock setting, click
Get Real Time Clock
.
3.
To send the computer’s clock settings to the arm, click
Set Real Time Clock
.
4.
Click
OK
.