Chapter 11: Configuring the Arm Using the Prosthetist Interface
2.
Click
OK
to lose changes to current configuration.
3.
Click
Update Arm
to load a cleared, blank configuration to the arm.
The Save Configuration to File screen opens. See
.
Figure 118. Save Config to File
4.
Click
Yes
if you would like
to save a copy of the cleared, blank configuration for
future reference.
5.
Click No to continue to update the arm without saving a copy of the
configuration.
6.
Power cycle the arm. See
.
Read the Current Configuration from the Arm
To read the current arm configuration perform the following steps:
1.
Click the
Read Configuration
button.
The current arm configuration is read to the system.
Note that at this point, the configuration is cleared from the computer only
and is not cleared from the arm. Update Arm must be performed to send the
cleared configuration to the arm. Once the configuration is cleared from the
arm, the arm will not operate until a new configuration is loaded.