Step 4: Configure Inputs
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
193
2.
Select the corresponding motion for the correct IMU as indicated in the IMU
Setup screen.
The input level should hover or “zero” somewhere around the midpoint of the
scale. If it does not, perform the following to “re-zero” the IMUs:
a.
Have the client stand with the IMUs in the desired zero position (usually feet
flat on the floor).
b.
Press the
Activate/Zero IMUs
button. See
The full red bar, as shown in
, is an indication that the IMUs have
not been zeroed. After zeroing the IMUs the red bar will be displayed as shown
in
.
Table 24. IMU Setup - Fields and Buttons
Fields/Buttons
Description
Input Type
The type of input used to control the arm.
The selections on the screen are:
•
IMU
•
Rate Sensitive
•
Pattern Recognition
NOTE: Selections that are not available are grayed
out.
Activate/Zero IMUs
Activates and zeros the IMUs.
Independent
Allows output of only the intended command axes
with the greater signal intensity on an IMU.
Coupled
Allows for simultaneous output of both axes of an
IMU.
Walk Detect
Enables (active) or disables (inactive) walk detect. Walk
detect ignores commands from the IMU when the
client is walking.