Appendix B: Technical Specifications
Quality of Service Provisions
Interruption or corruption of communication between the Arm and
IMUs can lead to interruptions in arm motion. Interruption of
communication for more than 2 seconds may lead to the system
reverting to Standby mode. Interruption of communication for more
than 8 seconds results in the system declaring a fault.
Common consumer electronic devices that transmit in the same
frequency band used by the LUKE arm system may prevent the
Arm and IMUs from communicating. Microwave ovens, Bluetooth®
devices, Wi-Fi® networks and 2.4 GHz cordless phones, when
transmitting or receiving, can cause interruption of communication
between the Arm and IMUs. During testing, the LUKE arm system
experienced occasional communication interruptions in the
presence of Bluetooth mice. It is likely that other devices operating
in similar frequency ranges can have a similar effect. This
interference, however, will not cause any incorrect data to be sent
and will not cause any harm to the LUKE arm system.
Some metal detectors and anti-theft detection systems at store exits
transmit in the same frequency band used by the LUKE arm system.
These devices can cause interruption of communication between
the Arm and IMUs. Again, this interference will not cause any
incorrect data to be sent and will not cause any harm to the LUKE
arm system.
In each of these cases, communication problems can usually be
resolved by turning off or moving away from other RF transmitting
devices.
Table 46. Arm Radio Specifications (Continued)
Parameter
Range/Explanation