Troubleshooting the Arm
Rev 1.8 -- Use or Disclosure of Data Contained on this Page is Subject to the Restrictions on the Inside Title Page of this Document
12/5/16
103
Low Battery Icon On
Battery is low
Replace the external battery in the holster
with a fully charged battery.
Plug the AC Adapter into the forearm
charging port
External battery not
connected
1.
Check that the external battery is prop-
erly seated in the holster
2.
Check and tighten the cables between
the arm and the holster
Grip Select LEDs Sweeping
IMU not talking to arm
Shake the IMUs to wake them
IMU battery is low
Charge the IMUs
Ham radios, walkie
talkies, theft detectors, or
metal detectors are
affecting the arm
Move the arm at least 0.5 m (20 inches)
away from any ham radios, walkie talkies,
theft detectors, or metal detectors
Arm Mode LED Blinking
Walk Detect
1.
Stop walking
2.
Put your feet flat on the ground
3.
Check that the Arm Mode LED has
stopped blinking
IMU tilted too far
1.
Put the arm into Standby Mode
2.
Check and secure the IMUs on your
feet
3.
Put your feet flat on the ground
4.
Take the arm out of Standby Mode
Input device not working
Contact Technical Support. See
Table 19. Troubleshooting - Wrist Display and System Faults
Problem
Cause
Solution
Sweeping