![Mobius Bionics LUKE Arm Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/mobius-bionics/luke-arm/luke-arm_user-manual_1815915048.webp)
108
Use or Disclosure of Data Contained on this Page is Subject to the Restrictions on the Inside Title Page of this Document -- Rev 1.8
Appendix A: Technical Specifications
:
:
IMU IP Rating
IP57
Table 23. Operating Environmental Range (Continued)
Parameter
Range/Explanation
When using the arm continuously in a hot environment (40°C, 104°F) and
while charging the internal battery, portions of the forearm and upper arm (if
applicable) could reach temperatures of 54°C - 57°C (130°F - 135°F), when
evaluated as directed in IEC60601-1: 2005-12.
Table 24. Transport and Storage Environmental Range
Parameter
Range
Storage Temperature (excluding
battery)
-25
°C
to 70
°C
(-13
°F
to 158
°F
)
Humidity
15% to 93% (non-condensing)
Pressure
700 hPa to 1060 hPa
Table 25. Service Life Specifications
Part
Service Life
Arm
Expected to function for up to three (3) years with an 18 month
service interval
Internal and External Batteries
Expected to provide at least 80% of new capacity for up to a year of
typical use
IMU Battery
Expected runtime is at least 18 hours for up to a year of use
AC Adapter and Charging Pad
Expected to function for up to three (3) years
External Battery Holster and
External Battery Charging Dock
Expected to function for up to three (3) years
ACI and Cables
Expected to function for up to three (3) years