![Mobius Bionics LUKE Arm Скачать руководство пользователя страница 139](http://html1.mh-extra.com/html/mobius-bionics/luke-arm/luke-arm_user-manual_1815915139.webp)
Step 4: Configure Inputs
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
195
b.
Setting the Medial and Lateral Thresholds such that Anterior and Posterior
movements do not cause the Medial/Lateral signal level to cross the
threshold and cause unintended motions.
c.
Setting the Thresholds by entering the threshold value in the threshold box
or by dragging the corresponding threshold sliders.
6.
Repeat Steps 4 and 5 for the Right IMU.
7.
If the client is unable to independently actuate each motion of the IMU with
modest thresholds, the independent option for the IMU axes should be selected.
This allows for command of one axis at a time. If the independent option is
selected repeat steps 1 through 7 above.
8.
Click the
Update Arm
button.
If, after configuring the action (
STEP 5
in the configuration process), the arm
moves due to crosstalk of the pitch and roll commands, check that the IMUs
are properly positioned then increase the Threshold until the unintended
motion stops. As an alternative, selecting the independent option allows for
command of only one axis at a time.