Troubleshooting the Arm
Rev. 1.5 -- Use or Disclosure of Data Contained on this Page is Subject to the Copyright Restrictions on the Inside Title Page of this Document
229
Low Battery Icon On
Battery is low
1.
Replace the external battery in the hol-
ster with a fully charged battery.
2.
Plug the AC Adapter into the forearm
charging port
External battery not
connected
1.
Check that the external battery is prop-
erly seated in the holster
2.
Check and tighten the cables between
the arm and the holster, and replace
any damaged cables if necessary
Grip Select LEDs Sweeping
IMU not talking to arm
Shake the IMUs to wake them
IMU battery is low
Charge the IMUs
Arm trying to talk to
wrong set of IMUs (if two
sets are configured)
While the LEDs are still sweeping, press the
wrist display button
Ham radios, walkie
talkies, theft detectors, or
metal detectors are
affecting the arm
Move the arm at least 0.5 m (20 inches)
away from any ham radios, walkie talkies,
theft detectors, or metal detectors
Table 30. Troubleshooting - Wrist Display and System Faults
Problem
Cause
Solution
Sweeping