CHAPTER 12: MAINTENANCE
LMV1000 (G)(E) AIR-COOLED LASER MARKER
990-541
12-9
18.
Replace the LD insulating cover.
19.
Install the top cover back onto the
Control Unit. And slide the top cover
forward to secure it.
20.
When starting the equipment for the
first time after an LD Module change,
be sure to reset the LD total hours,
wavelength and LD current limit values
(see next section). Failure to update this
information may result in decreased
efficiency and decreased output power.
Resetting/Changing the LD Module system parameters
The LD Module will degrade over its lifetime.
To compensate for the degradation, the LD
current should be raised approximately 1A for
every 1000 hours of use. The LD Current limit
may be raised as necessary to fulfill this
requirement.
Whenever an LD Module is changed, the
following parameters
must
be updated in the
system:
x
LD Hours
x
LD Wavelength
x
LD Current limit
NOTE:
Refer to
Chapter 6, Operation With the
Touch Panel, (7) Maintenance Screen
and the
supplied LD Data Sheet for details.
Содержание LMV1000
Страница 8: ...LMV1000 G E AIR COOLED LASER MARKER viii 990 541...
Страница 9: ...LMV1000 G E AIR COOLED LASER MARKER 990 541 ix...
Страница 17: ......
Страница 35: ......
Страница 47: ......
Страница 215: ...8 COMMUNICATION FUNCTION OPTION LMV1000 G E AIR COOLED LASER MARKER 8 14 990 541...
Страница 231: ......
Страница 237: ......
Страница 238: ...LMV1000 G E AIR COOLED LASER MARKER 990 541 11 1 CHAPTER 11 OUTLINE DIMENSIONS 1 Dimensions Control Unit Oscillator...
Страница 239: ...11 OUTLINE DIMENSIONS LMV1000 G E AIR COOLED LASER MARKER 11 2 990 541 Oscillator Continued...
Страница 241: ...11 OUTLINE DIMENSIONS LMV1000 G E AIR COOLED LASER MARKER 11 4 990 541 Oscillator 1 2 3 4...
Страница 243: ......
Страница 259: ......