Appendix A
DVVA II 4000
A-24
400-42803-01 (Rev B) - July 2017
6051-
Analysis cancelled. The difference between upper and lower load cell readings is
too high. Check that the sample tray is in place.
6052-
Analysis cancelled. The upper piston has reached the minimum allowed height.
Check that the sample tray is in place.
Cause B:
The lower piston drive assembly is damaged.
Action B:
Contact your service representative.
Cause A:
The sample tray containing the lower piston was not in place when the
analysis was started.
Action A:
Lower the lower piston, open the tray drawer, and install the sample tray.
Cause B:
The upper or lower piston does not have a proper calibration.
Action B:
Contact your service representative.
Cause A:
The sample tray containing the lower piston was not in place when the
analysis was started.
Action A:
Lower the lower piston, open the tray drawer, and install the sample tray.
Cause B:
The linear transducer does not have a proper calibration.
Action B:
Contact your service representative.