Appendix A
DVVA II 4000
A-20
400-42803-01 (Rev B) - July 2017
6025- Unable to reach the requested ending pressure.
6026-
The instrument is currently disabled. Ensure that the shield is down and the
drawer closed, and press the instrument reset button.
6027-
Instrument fault.
6028-
The sample file has an invalid status and cannot be used for analysis.
6029-
Application restarted during an analysis. Analysis canceled.
Cause:
The instrument could not build pressure to the endpoint specified in the
Analysis Conditions dialog.
Action:
Check that the ending pressure in the analysis conditions is within the
specification for your instrument. If it is, then call your service repre-
sentative.
Cause:
One of the safety interlocks was triggered.
Action A:
Ensure that the safety shield for the sample compartment is completely
down, and that the disposal drawer is completely closed. Then press the
Reset
button.
Action B:
Turn the red
Emergency Stop
button clockwise to reset it, then press the
Reset
button.
Cause:
You selected a sample file with a status other than
No Analysis
.
Action:
Select a sample file that has not been used in an analysis.
Cause:
You exited the application during an analysis and when you restarted the
application, the analysis had been canceled.
Action:
None required. The data collected up to the point of cancellation will be
saved in the sample file.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......