Appendix A
400-42803-01 (Rev B) - July 2017
A-5
2456-
Insufficient file handles available. Application cannot continue.
2457-
Results cannot be displayed. More than (number) windows are currently displaying
or printing results.
2458-
An instrument is performing a critical operation. Wait a few moments before exiting
the application.
2459-
An instrument is busy. A delay in restarting this application could result in loss of
new data. Continue with program Exit?
Cause:
You have more than 50 files open at the same time.
Action:
Refer to the manual for your operating system and set the limit for
open files to 50 or greater.
Cause:
You have too many windows open in the application.
Action:
Close some of the open windows.
Cause:
You attempted to exit the application while the analyzer is performing
a critical operation. This operation must be completed before the appli-
cation can be stopped.
Action:
Wait a few minutes before attempting to exit the application again.
Cause:
You attempted to exit the application while the analyzer was perform-
ing an operation.
Action:
Rather than exiting the application you may choose to minimize it.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......