6039-
Data Acquisition error.
6041-
Not enough data collected in file (file name) for reporting.
6044-
Use Prior Blank selected, but a prior blank could not be found.
6050- The lower piston did not successfully set the home position.
Cause A:
Network activity is interfering with communication.
Action A:
Close programs that use the network and disconnect and shared drives.
Cause B:
You had other applications open while analysis data were being col-
lected.
Action B:
Close all other applications when performing an analysis.
Cause C:
A scheduled update was being performed while analysis data were being
collected.
Action C:
Schedule automatic updates during a time when analyses are not being
performed; for example, during non-working hours.
Cause D:
The hardware configuration of the computer was changed during anal-
ysis.
Action D:
Do not add or remove devices such as USB drives during analysis.
Cause:
You requested a report for a file that does not contain collected data.
Action:
Choose a file that contains collected data (
Complete
status).
Cause A:
There was an obstruction in the lower piston path when the tray loading
position was being set.
Action A:
Open the tray drawer and repeat the sample tray.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......