DVVA II 4000
Normalization
400-42803-01 (Rev B) - July 2017
9-29
6. Using the table below, enter in the
Bulk Density
field the bulk density of the reference
material for this run.
7. Click
Start
to begin the blank run.
8. When the blank run has finished, the following dialog is displayed:
9. Perform the following steps:
a. Open the sample shield.
b. Pour the reference material into the sample chamber.
c. Wipe the piston tip with a Kimwipe or paper towel.
d. Close the sample shield and press
Reset
.
e. Click
OK
on the Load Sample dialog to begin the reference run.
10. When the analysis has completed, the
Start
push button on the dialog changes to
Next
; Click
Next
to return to the original view of the Analysis dialog.
Bulk Density for
SRB Reference Materials
Reference Name
Bulk Density
A-8 Tread/Hard
0.41
B-8 Tread/Hard
0.31
C-8 Tread/Hard
0.29
D-8 Carcass/Soft
0.64
E-8 Carcass/Soft
0.44
F-8 Carcass/Soft
0.34