Maintenance
DVVA II 4000
9-26
400-42803-01 (Rev B) - July 2017
11. Click
OK
; the report window containing overlay results is displayed. Click the
Height vs
Pressure
tab to display the data.
12. Right-click in the graph area and choose
Autoscale
from the shortcut menu.
13. Zoom in on these portions of the graph: 5,000, 15,000 and 25,000 psi. Continue zooming on
each until you can determine if the curves for three sequential runs are within 0.001 inch.
If the curves are not within this range, repeat analyses until three sequential curves are
obtained. If after a total of 20 analyses you still have not obtained these results:
• Refer to
Table 9-1. Common Operational Problems
, page
9-1
for corrective solutions.
• Contact your service representative if the solutions in Table 9-1 do not correct the
problem.