Appendix A
DVVA II 4000
A-16
400-42803-01 (Rev B) - July 2017
6003- File (file name) cannot be opened. It is currently being setup for an analysis.
6004
An analysis cannot be performed on (file name). It is open for editing and contains
errors.
6005- The edit session for (file name) must be saved before the analysis. Save changes
and continue with the analysis?
6006-
At least one report item must be selected. Press Cancel if you do not want a report.
Cause:
You attempted to open a file that is currently being used in an analysis.
Action:
Select a different file.
Cause:
You attempted to start an analysis using a file that contains errors and is
open for editing.
Action A:
Finish editing the file, save, and close it, or choose a different file. Then
restart the analysis.
Cause:
You attempted to start an analysis using a file that contains unsaved
changes and is still open.
Action:
Click
Yes
to save the changes and proceed with the analysis.
Click
No
to cancel the analysis and (1) continue editing the file, save and
close it; or (2) click
Browse
and choose a different file.
Cause:
You failed to select reports in the
Selected Reports
list window. No out-
put could be produced.
Action:
Ensure that at least one report is selected for the sample and re-submit
the sample for reporting, or click
Cancel
.
Содержание DVVA II 4000
Страница 1: ...DVVA II 4000 Dynamic Void Volume Analyzer Operator s Manual Rev B 400 42803 01 July 2017 ...
Страница 4: ......
Страница 10: ...Table of Contents DV VA II 4000 vi 400 42803 01 Rev B July 2017 ...
Страница 16: ......
Страница 32: ......
Страница 116: ......
Страница 136: ......
Страница 160: ......
Страница 180: ......
Страница 212: ......
Страница 214: ......
Страница 240: ......
Страница 246: ......
Страница 248: ......